Breathing circuits

Non-rebreathing circuits

Best suited for small animals under 7 kg. These systems require high fresh gas flow rates 100-300 ml/kg/min to remove exhaled gases. These systems will work partial rebreathing at or below gas flow rate of 130 ml/kg/min

Rebreathing systems

  • In NONrebreathing systems, there is NO remixing of inhaled and exhaled gasses. These systems run at HIGH fresh gas flow rates, 100-300 ml/kg/min.
  • A closed anesthetic rebreathing system only provides enough fresh gas flow to meet an animal’s metabolic needs, about 5-10 ml/kg/min (depending on animal size. Flow is lower for larger animals and higher for smaller animals).
  • Semi-closed and partial rebreathing systems are the same thing. They run at intermediate flow rates where fresh gas is delivered in excess of metabolic consumption, from about 10 ml/kg/min (large animals) or 30 ml/kg/min (small animals).

Bain system

  • A Bain system is a tube within a tube.
    • New oxygen and anesthetic gas is inhaled down the inner tube, and exhaled gas exits through the outer tube. (Note there is a modified form of a circular system called a “universal F circuit” that is also a tube inside a tube).
    • A Bain anesthetic circuit run at a HIGH flow rate of ~300 ml/kg/min will not allow rebreathing of exhaled gasses.
    • At MODERATE flow rates (ie: ~20-300 ml/kg/min or less) the Bain circuit functions as a PARTIAL rebreathing system, and the animal rebreathes some of the exhaled gasses.

Gas supply

  • Gases and Color codes:
    • No matter what color the tank is, ALWAYS read the tag or label on the tank first to verify the gas contents inside. One instructor in Surgical Nursing and Anesthesia has reported seeing of a green tank containing carbon dioxide instead of oxygen!
    • That being said, oxygen tanks in the U.S are typically white or green. Nitrous oxide tanks, (laughing gas) are blue. (Try remembering-”Laughing gas chases the blues”)
    • STRATEGY HINT: This is an “inclusion” question. That is, the choices “White or green” and ” Red or brown” each INCLUDE more possibilities of being correct than the single-color choices of blue or yellow.
      • If you were completely lost and had nothing else to go on, you might increase your chances of guessing right by choosing between the two inclusive choices of “White or green” and ” Red or brown”.
      • This WILL NOT guarantee you always get a question right, but if you are lost, use whatever you can, including strategies like this.
    • Refs: Bassert and Thomas, McCurnin’s Clinical Textbook for Veterinary Technicians, 8th ed. p. 1091.

Ventilation

Inverse ratio ventilation

Inverse ratio ventilation (IRV) is an alternative strategy for mechanical ventilation that reverses the classical inspiratory/expiratory scheme. This is achieved by modifying the inspiratory to expiratory (I:E) ratio, typically with the intention to increase oxygenation by increasing the mean airway pressure (MAP). Discussion of IRV requires an understanding of basic ventilator management which can be reviewed in a separate article. Here we discuss additional terms necessary to the utilization of IRV.

I:E Ratio

The I:E ratio denotes the proportions of each breath cycle devoted to the inspiratory and expiratory phases. The duration of each phase will depend on this ratio in conjunction with the overall respiratory rate. The total time of a respiratory cycle is determined by dividing 60 seconds by the respiratory rate. Inspiratory time and expiratory time are then determined by portioning the respiratory cycle based on the set ratio. For instance, a patient with a respiratory rate of 10 breaths per minute will have a breath cycle lasting 6 seconds. A typical I:E ratio for most situations would be 1:2, if we apply this ratio to the patient above, the 6-second breath cycle will break down to 2 seconds of inspiration and 4 seconds of expiration. Increasing the I:E ratio to 1:3 will result in 1.5 seconds of inspiration and 4.5 seconds of expiration. Thus, a “higher” I:E ratio results in less inspiratory time and more expiratory time in the same length of the breath cycle.

Source: Inverse ratio ventilation, NCBI Bookshelf