Key points

  • (mA, focal spot, SID, kVp) to reduce time to avoid motion blur.
  • mAs (tube current and exposure time product): mA x t = mAs mAs is a measure of radiation produced (milliamperage) over a set amount of time (seconds) via an x-ray tube
  • The reciprocity law- reaction of a photogenic emulsion to light will be equal to the products of the intensity of that light and the time of the exposure
    • 50 mA x 1/5 sec. = 10 mAs is equal to 300 mA x 1/30 sec. = 10 mAs
  • kVps: Kilovoltage peak (kVp) is the peak potential applied to the x-ray tube, which accelerates electrons from the cathode to the anode in radiography or computed tomography.
  • DR System:
    • image quality parameters such as detective quantum efficiency. Detective quantum efficiency (DQE) is one of the fundamental physical variables related to image quality in radiography and refers to the efficiency of a detector in converting incident x-ray energy into an image signal. DQE is a measure of the combined effects of the signal and noise performance of an imaging system.
  • CR image under sunlight: photostimulable phosphor CR systems. Exposure to a bright light (including sunlight) fully de-excites the trapped electrons, erasing the stored image. The imaging plates are reusable thousands of times but are subject to physical wear and tear. The image is created on reusable storage phosphor imaging plates. when phosphor plates are exposed to x-rays, part of the radiation energy is absorbed by electrons, which store the image temporarily.
  • PACS: The term picture archiving and communication system (PACS) applies to the combined hardware and software used for digital imaging.
  • Techniques Charts: Veterinarians must develop new technique charts for their new CR system based on the manufacturer’s guidelines.
  • kVp Kilovoltage potential:
    • Determines the energy of xrays. High kVp techniques are most useful for studies of body regions with many different tissue densities (eg, thorax).
  • mA milliamperes: linear function, high mA increases the no. of photons/xrays. the amount of darkening on the image is related to the total number of photons reaching it. Therefore, increasing mA increases image contrast. Changes in mA settings are relatively linear; increased contrast is desirable when tissue densities are similar (eg, soft-tissue components of the musculoskeletal system). However, increasing mA generally results in more heat loading on x-ray tubes, thus limiting exposure times and reducing tube life.
  • Exposure time (s):
    • mAs = milliamperes (mA) x exposure time (s)
    • Increasing the exposure time increases the number of photons produced and hence the darkness of the image. For exposures in the general diagnostic range, this is a linear function; as is the case with increasing mA, increasing exposure time generally results in greater heat loading of the x-ray tube than increasing kVp, once again potentially shortening tube life.
  • All three of the above parameters are interdependent, with exposure time and mA so much so that the term milliampere-seconds (mAs) is usually used to indicate the product of these two factors. Increasing the mA and decreasing the exposure time by a proportionate amount results in a radiograph less likely to be degraded by motion. As a rule, it is best to minimize the exposure time but maintain an appropriate mAs and scale of contrast. Increasing kVp increases the number of photons penetrating the subject and so darkens the image. This effect can be used within limits to correct an underexposure. The converse is likewise true.
  • When correcting a previously unsatisfactory image, underexposure or overexposure should be corrected for by adjusting the mAs when examining areas of high contrast (skeleton) or by adjusting the kVp when examining areas of low contrast (thorax). This will maintain the same relative contrast for that anatomic area while adjusting the film darkness.
  • A technique chart must be made for each machine.
  • Radiopaque tissues/objects appear more white and radiolucent tissues/objects appear more black. The resultant pattern of opacities forms an image on the radiograph, which is recognisable in form, and which can be interpreted.

Principles of x-rays

When a radiographic image is obtained, about 90% of the x-ray photons are absorbed by the tissue and 10% of the photons pass through the patient and reach the detector. Many of the absorbed photons generate scattered radiation (Compton scatter). These scattered photons travel in all directions creating noise and degrading image quality. The effect of scattered radiation can be minimized by collimating the x-ray beam to reduce the number scattered photons.

The farther away you are from the patient the less intense the photons are that strike plate. This is especially important for obtaining radiographic images in large body parts with portable generators. The typical film focal distance (distance from the plate to the generator) is 60 cm. When imaging larger body parts such as the caudocranial stifle, neck or back it is important to maintain a 60-cm distance (or closer). We can also stand closer to the image detector and patient to improve image quality when imaging larger body parts when limited by our generators.

By standing closer when obtaining radiographs of larger body parts, the increased quantity and intensity of photons reaching the plate will help to reduce the commonly obtained gray, grainy images (also known as quantum mottle).

Source: Field Guide to Improving Radiographs, K. Selberg (VIN)

Radiographic opacity

Radiographic opacity refers to the actual penetrative ability of x-rays to pass through an object and reach the film. Radiographic opacity of a part is determined by its thickness and its atomic weight. Therefore, atomic weight and thickness are closely related.

Image formation is dependent on the phenomenon of differential absorption. When x-rays penetrate tissue, they are not homogeneously absorbed; some tissues absorb x-rays more efficiently than others. If x-ray absorption were uniform, the resulting radiographic image would be grey or white. Some of the x-rays are absorbed by the tissues such as bone. Other x-rays pass through the tissues and produce the diagnostic image on the film. Other x-rays pass into the tissues and are deflected or scattered in the tissues and may exit onto the film. These are unrepresentative of the tissue through which they have passed. These are scattered x-rays and cause distortion of the final image. These scattered rays also contribute to the radiation hazard of the procedure by virtue of their unpredictable exit from the animal.

At this point we must differentiate between different types of subject tissue density.

The term tissue density is used to describe the degree to which a patient or object absorbs incident x-rays. In the accompanying radiograph the bone tissue is denser than the adjacent soft tissue, but the differences in image tone should be described in terms of radiolucency and radiopacity. For example, in this radiograph, the soft tissues are more radiolucent than the bones.

Air and fat absorb relatively less radiation and are consequently radiolucent, so their images on the film will appear black and pale grey respectively. Bone and metal absorb much more X-radiation and are radiopaque, so their images are white. Most soft tissues in the body are composed mainly of water and appear as shades of grey. The radiopacity of most fluids (blood, urine, transudates, exudates, bile and cerebrospinal fluid) and non-mineralised non-adipose tissues (muscle, cartilage, tendons, ligaments, fascia and parenchymatous organs) is the same. Lead and other metals have high physical density and effective atomic number, which renders them extremely radiopaque.

Source: not recorded in the original note.

Patient and x-ray positioning

The ideal conditions for obtaining most radiographs of equine joints is with the horse standing square, except for non-weight-bearing views. Unfortunately, not all of our equine patients are willing to stand square and the equine practitioner must accommodate the patient to make quality radiographs. Radiographic examination of the stifle, especially the caudocranial image, is a good example of this principle. For example, the optimal angle for the caudocranial image is 10 degrees from caudal proximal to cranial distal with the horse standing square (the tuber calcaneus is in line with the tuber ischii). However, if the limb is slightly behind the vertical (camped out), or under the horse (camped under), the angle must be adjusted to accommodate the stance of the horse. If the horse is standing with the limb behind it, the angle (caudoproximal to craniodistal) will increase (be steeper) and vice versa for the horse that is stood under itself. When evaluating the caudocranial stifle image for adequate positioning, the tibial plateau should superimpose itself from cranial to caudal and the tibial tuberosity should be distal (10–15 mm) to the tibial plateau. If the tibial tuberosity is proximal to the tibial plateau, the angle is typically too steep. The x-ray generator should be centered about 8–10 cm proximal to the indentation created by the distal aspect of the thigh musculature as it transitions to the proximal crus area. Joint narrowing in the stifle may be masked or falsely created by inadequate positioning.

The most common areas of pathologic change in the stifle are associated with the medial femoral condyle and the lateral trochlear ridge,2,3 the caudo 45° lateral–craniomedial oblique highlights these areas well. A well-positioned Cd45L–CrM oblique should project the medial femoral condyle caudal to the tibial eminence and show the joint clearly. The superimposition of the medial femoral condyle with the tibial eminence can mask subtle concave defects. Just as with the caudocranial projection of the stifle, the angle of the x-ray generator should be 5–10 degrees proximal to distal in a square standing horse. The most common mistakes are to is be at too steep of an angle or being too lateral.

Limb positioning also affects acquisition of the lateral radiographic projection. If the horse is standing base wide the x-ray generator will have to be angled distally; if base narrow, the generator angle will be slightly proximal. Judging placement of the x-ray generator in a cranial to caudal fashion is best done by lining up with the heel bulbs and tarsus. The most common mistake is being too far cranial.

Radiographic examination of the neck has seen a dramatic increase in frequency in recent years. Unlike the distal extremities, the radiographer cannot see the x-ray detector on the opposite side of the neck. This creates positioning problems and often results in images with one vertebral body or facet joint that is not centered on the x-ray detector. Furthermore, the articular facets on the lateral radiographs may not be perfectly superimposed, which can lead to interpretation errors. However, if a moment is taken to palpate the transverse processes and apply white tape to these sites, this can serve as a guide to detector placement and x-ray generator focus. Centering just proximal to the transverse process will render well-positioned radiographs, when the horse’s poll is in line with the withers.

Symmetrical anatomy of the neck can make lesions difficult to lateralize. Oblique radiographs obtained in a left/right 45–55° ventral to right/left dorsal fashion can help localize lesions.5 The x-ray generator is typically centered at the jugular furrow and the x-ray detector is placed with the transverse process centered at the bottom ⅓ of the x-ray detector. Radiographic images are named from where the x-ray generator is located and subsequently where the x-rays enter to where the x-rays exit the neck, and where the x-ray detector is located. In the example of a L55V–RD oblique the left articular facets will be projected dorsally and the right transverse processes will be projected ventrally. Properly labeled, opposite oblique radiographs should be obtained to accurately localize the lesion. Well positioned oblique radiographs should project one side of the articular facets dorsal and show the intervertebral foramen well. The other articular facet joint will be superimposed over the vertebral body highlighting the joint width. Oblique radiographs obtained with portable units centered at the articular facets at C6–7 are challenging due the shoulder superimposition. This may be overcome by offsetting the forelimbs, with the leg near the x-ray generator pulled caudally.

Radiographic projections are used to highlight specific areas and document pathologic change. The fetlock is an area with a multitude of pathologic change in which appropriately positioned radiographs make the pathologic change easy to identify. For example, palmar/plantar process osteochondral fragmentation is a common abnormality in the fetlock and depending on the location may be a source of lameness. The oblique views (dorso 20° proximo 45° lateral–palmaromedial oblique and dorso 20° proximo 45° medial–palmarolateral oblique) should show these lesions best. However, if the proximal sesamoid bones superimpose this area, the fragmentation could be easily missed. Ideal oblique and DP radiographs of the fetlock project the proximal sesamoid bones proximal to the joint margin.

Source: Field Guide to Improving Radiographs, K. Selberg (VIN)

Bone changes on radiographs

Bone remodelling is an active and dynamic process that relies on the correct balance between bone resorption by osteoclasts and bone deposition by osteoblasts. The remodeling cycle consists of three consecutive phases: resorption, during which osteoclasts digest old bone; reversal, when mononuclear cells appear on the bone surface; and formation, when osteoblasts lay down new bone until the resorbed bone is completely replaced.

  • Enthesophyte formation is new bone at the site of attachment of a tendon, ligament, or joint capsule to bone.
  • Osteoporosis defines a group of skeletal disorders that are characterized by loss of bone mass (osteopenia) with the remaining bone appearing normal.
    • There is P3 osteopenia and remodeling of the distal aspect of P3.
    • Pathogenesis: Osteoporosis results from bone resorption exceeding bone formation.
    • Note: When inadequate bone mineralization is microscopically evident, the disease should not be called osteoporosis.
    • Causes: Disuse osteoporosis is caused by lack of physical exercise, reduction in weight bearing and immobilization of the limbs. Osteoporosis of senility may be partially due to reduced physical activity. Malnutrition, hyperadrenocorticism, prolonged use of corticosteroids, thyrotoxicosis, hyperparathyroidism and other hormone imbalances also can cause osteoporosis.
  • subchondral sclerosis?
    • A sclerotic lesion is an unusual hardening or thickening of your bone.
    • Remodeling imbalance owing to failure of the resorptive process can result in dense (sclerotic) bones. Bone sclerosis is defined as “an abnormal increase in density and hardening of bone”
  • Ankylosis is a stiffness of a joint due to abnormal adhesion and rigidity of the bones of the joint, which may be the result of injury or disease.
  • subchondral erosions
    • The formation of large “holes” in the body of the navicular bone itself, believed to be enlarged vascular channels

Notes

  • There is an indistinct axial margin with underlying ill-defined lucent zones, involving the proximal half of the medial and lateral proximal sesamoid bones (